JINLING-840 Anesthesia Machine

● Flow rate: O2 0.05 ~ 1 L/min; 1 ~ 10 L/min    N2O 0.05 ~ 1 L/min; 1 ~ 10 L/min

Air 0.2 ~ 1 L/min; 1 ~ 12 L/min   ORC O2 ≥ 25%

● O2 flush: 35 ~ 75 L/min

● Insufficient oxygen alarm and automatic N2O calibration function:

When O2 level is ≤ 0.2 Mpa, audible alarm sounds and lasts for at least 7 seconds

When O2 level is ≤ 0.14 Mpa, N2O flow decreases correspondingly

When O2 level is ≤ 0.04 Mpa, N2O flow is cut off completely.

Product Description

Specifications:

● Flow rate: O2 0.05 ~ 1 L/min; 1 ~ 10 L/min    N2O 0.05 ~ 1 L/min; 1 ~ 10 L/min

Air 0.2 ~ 1 L/min; 1 ~ 12 L/min   ORC O2 ≥ 25%

● O2 flush: 35 ~ 75 L/min

● Insufficient oxygen alarm and automatic N2O calibration function:

When O2 level is ≤ 0.2 Mpa, audible alarm sounds and lasts for at least 7 seconds

When O2 level is ≤ 0.14 Mpa, N2O flow decreases correspondingly

When O2 level is ≤ 0.04 Mpa, N2O flow is cut off completely

● Integrated breathing circuit system, APL adjustment range: 0.5 ~ 7 kPa

● CO2 absorber capacity: 1.6 L

● Vaporizer ambient temperature: 15 ~ 35 °C

● Input flow: 0.25 ~ 15 L/min

● Output back pressure: -10 kPa ~ 20 kPa

● Output concentration: 0.2 % ~ 5% (± 0.2 %) (Vol %)

● Ventilation frequency: 6 ~ 40 bpm

● Tidal volume; adult: 100 ~ 1100 ml Infant: 30 ~ 150 ml

● I:E: ratio 1:1 ~ 1:9.9 (adjustable)